Máy Đo – Phân Tích Công Suất – Power Analyzer 4 Channel ZES LMG641

Thông tin sản phẩm

  • Hãng sản xuất: ZES Zimmer
  • Mã sản xuất: LMG641
  • Bảo hành: 24 tháng

Catalog

Duapath Explained

Data sheet sensor

Data sheet LMG test suite

Giá: Liên hệ

Thông tin liên hệ

Yêu cầu báo giá

Sự kết hợp kênh phù hợp cho mọi ứng dụng

Cấu hình với 1 đến 4 kênh đo công suất (thay thế 3 kênh đo và Giao diện tín hiệu xử lý)

Trang bị thêm các kênh có sẵn!

Inverter measurements
1-phase Input
3-phase Output
power measurement
of 3-phase motors
Efficiency measurement
of power supplies
Measuring of core losses

Measurement Channels

L60-CH-A A-Channel for high-precision broadband measurements
High-Precision Channel for power measurements with DualPath functionality, bandwidth DC-10 MHz
best accuracy 0.015% + 0.01%
Measuring inputs and ranges:
1. Voltage directly (V): 3/6/12,5/25/60/130/250/400/600/1000 Vtrms; 3200 Vpk
2. Voltage sensor input (mV): 30/60/120/250/500 (V): 1/2/4 Vtrms; 8 Vpk
3. Current directly (mA): 5/10/20/40/80/150/300/600 and (A): 1.2/2.5/5/10/20/32 Atrms; 120 Apk
4. Current sensor input (mV): 30/60/120/250/500 (V): 1/2/4 Vtrms; 12.5 Vpk
L60-CH-B B-Channel for general purpose measurements
Multi-purpose measurement channel with DualPath functionality, bandwidth DC – 500 kHz
best accuracy 0.05% + 0.02%
Measuring inputs and ranges
1. Voltage directly (V): 3/6/12.5/25/60/130/250/400/600/1000 Vtrms; 3200 Vpk
2. Current directly (mA): 5/10/20/40/80/150/300/600 and (A): 1.2/2.5/5/10/20/32 Atrms; 120 Apk
3. Current sensor input (mV): 30/60/120/250/500 (V): 1/2/4 Vtrms; 12.5 Vpk
L60-CH-C C-Channel for high-precision 50/60 Hz measurements
High-precision measurement channel for 50 Hz applications, bandwidth DC – 10 kHz,
best accuracy 0.03% + 0.01%.
Measuring inputs and ranges:
1. Voltage directly (V): 3/6/12,5/25/60/130/250/400/600/1000 Vtrms; 3200 Vpk
2. Current directly (mA): 5/10/20/40/80/150/300/600 and (A): 1.2/2.5/5/10/20/32 Atrms; 120 Apk
3. Current sensor input (mV): 30/60/120/250/500 (V): 1/2/4 Vtrms; 12.5 Vpk

Options

L6-OPT-DRV Option package for frequency converters
for LMG600 (at least 3 power channels*), consisting of:
– L6-PSI: Process Signal Interface
– L6-SDC: Star-Delta-Conversion
– L6-HRM: Harmonic analysis up to max. 1.000th order (2.000th order with interharmonics)
L6-OPT-TRAFO Option package for transformers
for LMG600 (at least 3 power channels*), consisting of:
– L6-SDC: Star-Delta-Conversion
– L6-HRM: Harmonic analysis up to max. 1.000th order (2.000th order with interharmonics)
L6-OPT-CE-FLK Option Package CE Flicker
for LMG600, consisting of:
– the LMG flicker meter according to EN61000-4-15 (L6-OPT-FLK) as well as
– the PC software** LMG Test Suite (LMG-TEST-CE-FLK) for analysis of measured values according to EN61000-3-3 and EN61000-3-11 for 1-phase and 3-phase* equipment under test
L6-OPT-CE-HRM Option Package CE Harmonics
for LMG600, consisting of:
– the harmonic analysis of the measured signals in the LMG (L6-OPT-HRM) as well as
– the PC software** LMG Test Suite (LMG-TEST-CE-HRM) for analysis of measured values according to EN61000-3-2 and EN61000-3-12 for 1-phase and 3-phase* equipment under test
L6-OPT-CE-HRMFLK Option Package CE Harmonics and Flicker
for LMG600, consisting of:
– the harmonic analysis of the measured signals in the LMG (L6-OPT-HRM),
– internal flicker meter according to EN61000-4-15 (L6-OPT-FLK) as well as
– the PC software** LMG Test Suite (LMG-TEST-CE-HRM + LMG-TEST-CE-FLK) for analysis of measured values according to EN61000-3-2 / EN61000-3-12 and EN61000-3-3 / EN61000-3-11, each for 1-phase and 3-phase* equipment under test
L6-OPT-DVI DVI / VGA Interface
for connection of an external display, beamer, touch device
L6-OPT-PSI Process Signal Interface
8 digital inputs, 8 digital outputs, 2 fast analog inputs (150 kS/s, 15 kHz bandwidth), 8 slow analog inputs (DC), 32 analog outputs, 2 frequency inputs (frequency, orientation, position; DTL, HTL or differential)
L6-OPT-CAN CAN bus interface: find more information here
CAN bus connectivity according to ISO 11898-2 (high-speed CAN)
– bit rate: 5 kbit/s to 1Mbit/s
– galvanic separation at CAN connector: up to 500V
– active support of CAN 2.0A and B standards
– Standard Frame Format or Extended Frame Format for
sending measurement results
– manual polling of measurement values possible via Remote Transmission Request (RTR)
L6-OPT-SDC Star-Delta-Conversion
Minimum requirement LMG600 with at least 3 power channels of the same type
L6-OPT-EVT Event Triggering: find more information here
triggering on crossing voltage or current thresholds recording of 16 tracks with up to 16 million values each
L6-OPT-HRM Harmonic analysis
up to max. 1.000th order (2.000th order with interharmonics) of U, I, P, Q and S, fundamental 3 … 1200 Hz, up to 150 kHz (A channel) resp. 15 kHz (B channel) or 10 kHz (C channel) with anti-aliasing filter, customizable settings for the detection of interharmonics
L6-OPT-FLK Flicker meter according to EN61000-4-15
L6-OPT-SMV Smart Vision
Smart add-on functions for the convenient evaluation of measurements
The package includes:
– Painter: screenshots can be enriched with valuable additional information directly in the GUI before saving
– Env Var Watcher: script-controlled visual watchdog as part of a custom menu
L6-OPT-SPV Sample Value Analysis Module for LMG600
Software* for continuous and gapless streaming of sample values on 16 tracks at a rate of 10 kS/s (up to 1.2 MS/s under optimal conditions), furthermore:
– display of sample recordings in ZLR format
– display of frequency spectrum of recorded sample values
– calculation of RMS values of voltage, current and power
– CSV or MATLAB export of frequency spectrum and RMS values
L6x1-OPT-MSD Mass Storage Device for Measurement Results
Internal mass storage device (SSD) for long-time recording of
measurements. Corresponds to a recording time of more than
600 days with a cycle time (cycle) of 100ms and 100 measurement values

* Note: power analyzer has to be equipped with at least three channels of same type for 3-phase measurements

** System requirements:
– Operating system: Windows 7/8/10 (32/64 bit)
– Hard drive: software: min. 50 MB, data: approx. 20 MB per minute of measurement/phase
– RAM memory: min. 2 GB
– Processor: min. 2 GHz, dual-core